ISOMED – SINGLE PHASE ISOLATION TRANSFORMERS for medical equipments supply from 0.3 to 2kVA
Isolation transformers for medical equipments have been designed to obtain the highest isolation from the electric network for the supply of medical devices, WITH AN EARTHED SCREEN between primary and secondary winding.
They are fitted with a 10A green led 2 poles circuit breaker on the front, while on the back side are placed all connectors, input IEC with 2 fuses, 4 or 8 output IEC sockets to EN60320-2 rated 10A each, output fuse protection. A proper connection cable is supplied. Metal case IP21, painted RAL7032 with lifting handle. Higher ratings available on request.
- installed in IP21 metal enclosure
- ral7035 epoxy painted
- lifting handles
- 2-poles illuminated 10A rocker switch on front
- input plug to EN60320-1 with 2-poles input fuse
- 2 nos. input time delay fuses according to the rating
- thermal resettable protection against short circuit, in series with primary winding
- isolation transformer with copper earthed screen
- 8 output sockets (4 on 300-500VA model) to EN60320-2 each max 10A (max current limited to max power of the unit)
- extra auxiliary equipotential earth male connector
- Output fuse according to rating
- Rating plate, operating manual, CE certificate
- 3000-4000VA models available (not complying with EN60601) with terminal blocks on input and output side, cables are to be provided at the place of the installation according to the requirements of the standards
| Rated power: | From 300VA to 2000VA |
| Input voltage: | 230V (220V) |
| Output voltage | 230V (220V) |
| Frequency: | 50/60 Hz |
| Degree of protection of the transformer | IP00 |
| Protection class: | I |
| Thermal insulation class: | F |
| Thermal class of the winding: | F-H |
| Ambient temperature: | 25°C |
| Max temperature rise: | 80°C |
| Max voltage drop: | 3,5% |
| Short circuit voltage: | 5% max |
| Full load efficiency | > 97% |
| Cooling | AN |
| Max relative humidity (non-condensing) | 95% |
| Operating altitude | up to 1000 m without derating |
| EEC Directives | Dir. 2014/35/EU – Dir. 2014/30/EU – Dir. MDR |
| Rules applied: | EN 61558-1, EN61558-2-4 |
| EN 60601-1:2006/AC:2010/A11:2011/A1:2013
Medical electrical equipment Part 1: General rules for safety |
The product is supplied as standard with a mechanical locking system of the outgoing plugs that prevents unwanted detachment if the cable is pulled.
The correct connection of the network cable is ensured by the lock system that blocks the mobile input socket to the panel plug that can only be unlocked by pressing the appropriate button.
The product can also be fixed from the bottom, being equipped with 4 reinforced threaded holes that allow, using M4 screws, to mount the product irremovably on a pre-drilled base, according to the fixing centers shown on the characteristics table
On the back side there is a POAG plug connector for earth connection